GSTDTAP  > 地球科学
DOI10.2172/1019542
报告编号DOE/SC-ARM/TR-072
来源IDOSTI ID: 1019542
CO (Carbon Monoxide Mixing Ratio System) Handbook
Cook, DR
2011-02-23
出版年2011
语种英语
国家美国
出版者PNNL;Richland, WA
领域地球科学
英文摘要The main function of the CO instrument is to provide continuous accurate measurements of carbon monoxide mixing ratio at the ARM SGP Central Facility (CF) 60-meter tower (36.607 °N, 97.489 °W, 314 meters above sea level). The essential feature of the control and data acquisition system is to record signals from a Thermo Electron 48C and periodically calibrate out zero and span drifts in the instrument using the combination of a CO scrubber and two concentrations of span gas (100 and 300 ppb CO in air). The system was deployed on May 25, 2005.